Procedure – Powder/Chip Samples
Determine the instrument blank.
Press the Analyze button on the instrument screen, the lower electrode will close and the analysis sequence will start and end automatically.
Repeat steps 2b through 2g a minimum of three times.
Set the blank following the procedure outlined in the operator's instruction manual.
Instrument calibration/drift correction.
Login a minimum of three Standard reps.
Weigh approximately *** grams of a calibration/drift standard into a *** Nickel Capsule; enter the mass and standard identification into appropriate rep fields. If applicable, place the nickel capsule in the appropriate autoloader position.
提供各種無機(jī)材料、鋼鐵、有色金屬、硬質(zhì)合金陶瓷材料中寬范圍的氧氮?dú)渫瑫r(shí)測(cè)定。新疆專業(yè)元素分析儀多少錢
Sample Preparation
Typically, titanium and zirconium samples are chemically etched to remove surface contamination when oxygen and nitrogen are determined. However, etching can introduce hydrogen into the sample.
ASTM method E 1409 "Determination of Oxygen and Nitrogen in Titanium and Titanium Alloys by the Inert
Gas Fusion Technique", as updated in 1996, permits either etching or abrading (filing) of the test specimen.
ASTM E 1447 "Determination of Hydrogen in Titanium and Titanium Alloys by the Inert Gas Fusion Thermal
Conductivity/Infrared Detection Method" permits surface preparation by abrading (if necessary to remove
contamination). Differences in sample preparation present somewhat of a dilemma regarding simultaneous
determination of O, N, and H in titanium. However, abrading samples with a file to remove surface
contamination will yield accurate O, N, and H results. The ONH836 utilizes a high-power electrode furnace to quickly and efficiently release the target gases from within the sample, which allows for a very rapid simultaneous determination of oxygen, nitrogen, and hydrogen.
江蘇LECO分析儀供應(yīng)商力可**的動(dòng)態(tài)流量補(bǔ)償控制DFC系統(tǒng),保證了流量系統(tǒng)的完善,可以精確控制整個(gè)氣路中的流量。力可OHN736&836系列儀器采用惰性熔融法,提供各種無機(jī)材料、鋼鐵、有色金屬、硬質(zhì)合金陶瓷材料中寬范圍的氧氮?dú)渫瑫r(shí)測(cè)定。ONH836儀器通過***采納客戶的反饋意見和創(chuàng)新設(shè)計(jì),力可***儀器采用觸摸屏技術(shù)和在線軟件平臺(tái),增強(qiáng)了儀器的可操作性,降低了分析成本,節(jié)省實(shí)驗(yàn)室桌面空間。
用戶友好界面的Cornerstone? 軟件
力可**的Cornerstone碳硫分析軟件采用觸摸屏界面使用戶可完整的達(dá)到分析控制、方法設(shè)定、在線診斷、數(shù)據(jù)報(bào)告。Cornerstone碳硫分析軟件允許用戶在單一分析界面中完成日常的所有工作,快速而且簡(jiǎn)單易用。力可創(chuàng)新的批次和重復(fù)樣品數(shù)據(jù)分組功能極大簡(jiǎn)化了樣品結(jié)果輸出和結(jié)果標(biāo)準(zhǔn)偏差的自動(dòng)計(jì)算。減少了繁瑣的額外數(shù)據(jù)處理。
Procedure – Powder/Chip Samples
Instrument calibration/drift correction.
Add approximately 0.05 g of *** Graphite Powder to a***Graphite Crucible.
Firmly place the crucible on the lower electrode tip or appropriate autoloader position.
Press the Analyze button on the instrument screen,the lower electrode will close and the analysis sequence will start and end automatically.
Repeat steps 3b through 3i a minimum of three times for each calibration/drift standard used.
無需人工干預(yù)的全自動(dòng)清掃系統(tǒng)。Software Features & Benefits
Cornerstone brand software is divided into four main sections—Analysis, Diagnostics, Settings, and Instrument—for simplified navigation and organization. Toolbars, sliders, and drop-down menus makeit easy to set parameters for calibration and data processing. The software also includes real-time ambient monitors, with fully animated system diagrams. Advanced diagnostic features include a thorough digital on-board manual, maintenance animations, photo illustrations, and screen captures that quickly provide the direction needed without having to refer to multiple manuals. also Cornerstone supports extended data archiving and flexible reporting capabilities. ONH836儀器通過***采納客戶的反饋意見和創(chuàng)新設(shè)計(jì)。浙江元素分析儀咨詢報(bào)價(jià)
可選六軸可旋轉(zhuǎn)觸摸屏幕,增強(qiáng)了操作的工效和直觀性。新疆專業(yè)元素分析儀多少錢
Procedure - Solid Samples:
--Instrument calibration/drift correction.
Place the nickel basket containing the calibration/drift standard into the open port at the top of the loading head.
Press the Analyze button on the instrument screen again, the loading head slide-block will close and the lower electrode will open.
Clean the upper and lower electrode manually, or, if applicable, remove the crucible and press the analyze button to clean with the automatic cleaner.